In-house Optical Pyrometer Calibration
Ease of Operation and Setup
Temperature Range: 1400°F - 4200°F
(800°C - 2300°C)
Calibration Accuracy to: ±5°F
Source: Tungsten Filament Strip Lamp
Certificate of Calibration to NIST
Power Requirements: 12 Vdc RV Marine
deep cycle battery
|
|
Calibration Test Set Datasheet
(81 KB)
|
Classification -
equipment calibration, optical pyrometer temperature calibration.

The Precision Optical Test Set calibration
standard is designed to accurately and simultaneously calibrate up to two optical pyrometers to read identically, irrespective of manufacture. It has been used for many years as a simple and easy-to-use
temperature calibration source to calibrate and verify optical pyrometers accuracy. In high demand applications, temperature errors and uncertainties are eliminated with on-site or in-house calibration.
The Optical Test Set uses a ribbon filament
tungsten strip lamp that is readily adjusted to the desired temperature value by adjusting the rheostat electrical current passing through the lamp. The lamp swivel base and sighting tube ensure precise optical alignment between the two optical pyrometers during calibration.


To operate the Optical Test Set, the user must first connect the proper power terminals to the 12 Vdc battery supply or equivalent DC power supply. Next, connect up a laboratory
precision dc ammeter or potentiometer with a corresponding shunt for 0 to 30 amperes to measure the electrical current going to the lamp. Finally, the operator uses the
filament lamp certificate of calibration data to determine the current necessary to produce the desired temperature of the lamp.

Model 107 Optical Test Set calibration
standard comes complete
with:
| |
- Lamp
Shield
-
Sighting
tube with mounted magnifying lens
-
Mounting
stem
- Vernier
slide
- Wire
rheostat (infinite resolution)
-
On-Off
temperature range switch
- Wiring
for battery charger
- DP-DT
switch for battery charger connection
-
Swivel
mount for Strip Lamp
-
Instrument holder for Optical pyrometer
-
Continuous step type105 points rheostat
-
Terminal
board and switch panel
- Ammeter
shunt switch
- Complete
installation and operating instructions
|
Requirements for the Model
107 Optical Test Set:

|
Model No.
|
Description
|
|
TS 41
TS 42
STL-L
STL-H
COC
P-907
|
EXTRA Instrument Holder (for
Optical pyrometer)
EXTRA Instrument Holder (for
Micro-Optical pyrometer)
STRIP LAMP Low
Temperature Range:
1400°-3400°F or 800°-1900°C
STRIP LAMP High
Temperature Range:
1400°-4200°F or 800°-2300°C
CERTIFICATION of CALIBRATION
(per point) traceable to
NIST
Digital Ammeter - for
measurement of current from
0 to 30 amps |

Good performance and reliability are what you expect when you buy precision products from The Pyrometer Instrument Company, Inc.. The Calibration Test Set
for optical pyrometer calibration comes complete with a one year factory warranty. For complete details visit our Product Warranty Page.
|